Currently viewing Vol. 1 • Issue 3 • 2014

A Hearing Aid Solution for Music

This page as PDF

Published on January 8, 2014. Reprinted with permission.

Figure 1. The average audiogram for the 10 subjects, with the standard deviations shown as the vertical bars at the respective audiometric frequencies.

Figure 1. The average audiogram for the 10 subjects, with the standard deviations shown as the vertical bars at the respective audiometric frequencies.

See if this sounds familiar: “I’ve had analog hearing aids for years and my banjo sounded great. They broke recently and I am now trying my fifth set of digital hearing aids. They distort the sound of my music and nothing that the audiologist does seems to help. I am not sure who is more frustrated—me or my audiologist!” I hear this type of comment frequently from musicians and many others. I also hear comments like this from many of my non-musician clients who, on occasion, like to attend concerts and generally like to listen to music.

The problem is not in the programming software. Tweaking the frequency response, gain, or compression features will only help if they were set up incorrectly for speech in the first place.

The problem is the analog-to-digital (A/D) converter that all digital hearing aids have. The A/D converters are found in the normal microphone route, the telecoil route, the direct audio input route, and (depending on the implementation) the wireless connection routes. The A/D converter is ubiquitous in modern hearing aid technology.

Partly because of the 16-bit architecture that modern digital hearing aids use, and partly because of some engineering design decisions that had to be made (generally to reduce the noise floor), modern hearing aids cannot handle overly intense inputs—this is typically the case for any input over about 95 dB SPL. This is true of entry level hearing aids and also true of top-of-the-line premium hearing aids. The most intense components of speech are on the order of the 80-85 dBA, so even shouted speech can get through the A/D converter “doorway” into the hearing aids.

The difficulty lies with music and other more intense speech inputs like a hearing aid wearer’s own voice. Even quiet music can have its peaks in excess of 95 dBA, and these music elements overdrive the A/D converter (also known colloquially as the “front end of the hearing aid”). Distortion that occurs as a result of A/D converters that are not up to the task of transducing music cannot be undone by software manipulation that occurs later in the hearing aid processing system. This front-end distortion does not respond to changes in frequency response, gain, or the hearing aid compression features that occur later in the hearing aid processing pathway.

The hearing aid industry has responded, in part, to this issue of poor music fidelity. A/D converter stages have been redesigned to “auto-range” or readjust their optimal region of digitization1-5 and also to utilize less sensitive microphones that “fool” the A/D converter into thinking that it is receiving lower level inputs that are better suited to their digitization parameters.6,7

A concern about these approaches is an increased noise floor in hearing aids that some people may hear—especially if their hearing sensitivity is good for the lower frequency regions. In addition, despite modern A/D converters having a theoretical limit of a 96 dB dynamic range (such as 7 dB SPL to 103 dB SPL), very few technologies have even been able to approach this limit.

Recently, a new approach has been introduced to the marketplace where the A/D converter dynamic range is elevated to an input region that is more appropriate for louder inputs such as music—like raising a low hanging bridge. True Input technology from Widex is the name associated with a novel approach designed to ensure that the available dynamic range is fully utilized while keeping the noise floor low. In addition, and perhaps most importantly, the technology utilizes a transformer action that allows the input voltage to be doubled. This has the effect of extending the top range of the A/D conversion process to 113 dB SPL, which is more appropriate for many forms of music.

Since the late 1980s, modern hearing aid microphones have been able to transduce inputs up to 115 dB SPL without distortion. With this new approach, full use is being made of the wide range of inputs that are available through the hearing aid microphone.

An Experiment

To study the effects of this new technology, 10 hard-of-hearing musicians who were experienced hearing aid users were asked to be assessed with, and wear, one of two hearing aids for a 6-week period: Widex Clear (without this True Input technology) and Widex Dream (with True Input technology). All subjects were fit binaurally and blinded to which hearing aids they were wearing. The researchers associated with the Musicians’ Clinics of Canada, who came into contact with the musician subjects, also were blinded with respect to which hearing aids were which. The settings and all other electroacoustic features of these two types of hearing aid were held constant, except for the different front end configuration of the Dream hearing aid.

There were two randomizations used for the 10 subjects: one with the Clear hearing aid being fit first, and the other with the Dream hearing aid being fit first. All 10 subjects were counterbalanced, and all 10 subjects had similar audiograms: mild-to-moderate high frequency sensorineural hearing loss with no more than a 40 dB HL hearing loss at 1000 Hz.

The data presented here are a subset of a larger study using other assessment tools. Figure 1 shows the average audiogram for the 10 subjects, with the standard deviations shown as the vertical bars at the respective audiometric frequencies.

All subjects were asked to listen to recordings under insert earphones of speech and music at low levels (-30 dB re: peak) and also at more intense levels (0 dB re: peak), and to make preference judgments of clarity and naturalness—attributes that have been shown to be useful in previous amplified music research.2 For these tests, the audiometer dial reading was adjusted for 85 dB HL, which corresponds with 102 dBA noise and a speech peak level of 108 dB SPL. The presentation level was adjusted to be “loud, but OK” for each subject.

At the end of the study, all hearing aids were returned along with the four weekly questionnaires. All subjects were paid for their participation according to the approved ethics protocol (University of Toronto, #27405) and were provided the opportunity to purchase the hearing aids at a discounted rate when the study was completed.

[caption id="attachment_751" align="aligncenter" width="100%"]Figure 3a. The same data as shown in Figure 2 but only for the music stimuli at both low and high presentation levels. Statistically significant differences (p<0.05) are denoted with an asterisk (*).Figure 3a. The same data as shown in Figure 2 but only for the music stimuli at both low and high presentation levels. Statistically significant differences (p<0.05) are denoted with an asterisk (*).

Figure 3b. The same data as shown in Figure 2 but only for the speech stimuli at both low and high presentation levels. Statistically significant differences (p<0.05) are denoted with an asterisk (*).

Figure 3b. The same data as shown in Figure 2 but only for the speech stimuli at both low and high presentation levels. Statistically significant differences (p<0.05) are denoted with an asterisk (*).

Results

Figure 2 shows the preference results in terms of clarity for the Dream and the Clear hearing aids for both high presentation level and low presentation level for both stimuli (ie, music and speech). At the high presentation level, the Dream circuitry performed statistically significantly better (p < 0.05) with a 50% difference over the Clear circuitry. There were no statistically significant differences for the lower level presentation. Similar data were also found for the subjective attribute of naturalness, but these are not shown due to space considerations. This provides evidence that the “raising the bridge” circuitry found in the Dream circuitry resulted in a preferred signal at the high input level, characteristic of many forms of music. The same data were then plotted separately for music and speech stimuli. Figure 3a shows the results for music, and Figure 3b for speech. The data were paired (a within-subject design) for each subject in order to minimize variance. For music, the Dream hearing aid (with True Input technology) was preferred statistically significantly more than the Clear hearing aid at both the high and the low presentation levels.

There was a 60% difference at the high input level and a 20% difference at the low input level. For speech, statistically significant differences were found at the high presentation level with the Dream circuitry performing better. There were no statistically significant differences at the lower presentation level.

Even for the quiet presentation level of music, there were statistically significant differences, and this is most likely related to the crest factor of the signal. The crest factor is the difference (in dB) between the instantaneous peak and the long term average root mean square (RMS) of the signal. For speech, the crest factor is on the order of 12 dB, whereas for music, the crest factor can be in excess of 20 dB.8 That is, the peaks for instrumental music are significantly higher with respect to the average RMS of the music than for speech that is presented at an equivalent level. Despite the lower presentation level of the music, the peaks most likely exceed the capability of the Clear circuitry, but not the newer Dream circuitry.

Speech is not typically uttered at such a high level as music, so one can argue that the result for the loud speech (Figure 3b) is superfluous; however, this condition is frequently encountered by hard of hearing people—their own voice at the level of their hearing aid microphones can be quite intense. The data for music in Figure 3a suggest that the “raising the bridge” circuitry of the Dream can be useful to improve the quality of a hard of hearing person’s own voice while wearing their hearing aids.

What This Means for Music and Musicians

Musicians, and those who like to listen to music, can now receive an amplified signal that is effectively distortion free. Software changes can still be made when necessary that can further optimize the listening environment, and now these changes can be made with as clear and high-fidelity of a signal as possible. The net result is a hearing aid fitting that is best suited to the requirements of playing, and listening to, music.

The data from this study indicate clearly that the “raising the A/D converter bridge” is a valid strategy that has been shown to be useful in both laboratory and real life listening environments. This is the approach that is emulated in the Widex Dream hearing aid with True Input technology. The concept of presenting the A/D converter in a hearing aid with a signal that is within its operating range (by designing the dynamic range of the A/D converter to be in a more appropriate location) has shown to minimize distortion and improve the sound quality, if the input level is high. In addition, this technology has ramifications of improving the quality of a hard of hearing person’s own voice.

Acknowledgements

The author is not associated with Widex but did receive compensation for the performance of the study.

The author thanks Francis Kuk and Brian Crose who assisted in the design of this study, its statistical analysis, and provided simulated materials for testing. He also thanks Anthony Vila and Brandie Beaudrie for assistance in coordinating the subjects.

References

  1. Killion MC. A high fidelity hearing aid. Hearing Instruments. 1990;41(8). www.etymotic.com.
  2. Chasin M, Russo FA. Hearing aids and music. Trends Amplif. 2004;8(2):35-47.
  3. Killion MC. What special hearing aid properties do performing musicians require? Hearing Review. 2009;16(2):20-31.
  4. Hockley NS, Bahlmann F, Chasin M. Programming hearing instruments to make live music more enjoyable. Hear Jour. 2010;63(9):30-38.
  5. Hockley NS, Bahlmann F, Fulton B. Analog to digital conversion to accommodate the dynamics of live music in hearing instruments. Trends Amplif. 2012;16:(3):146-158.
  6. Chasin M. Hearing aids for musicians. Hearing Review. 2006;13(3):11-16.
  7. Schmidt M. Musicians and hearing aid design—Is your hearing instrument being overworked? Trends Amplif. 2012;16(3):140-145.
  8. Chasin M, Hockley N. Some characteristics of Amplified Music through Hearing Aids. Hearing Research. 2013. July. Available at: http://www.sciencedirect.com/science/article/pii/S0378595513001743

Original citation for this article: A hearing aid solution for music. Hearing Review. 2014: January: 28-31

This page as PDF
About the author
Marshall Chasin, AuD

Marshall Chasin, AuD, Doctor of Audiology, Editor in Chief

Marshall is the director of research at the Musicians' Clinics of Canada and has presented and published extensively on the topics of hearing loss prevention in musicians and hearing aids for music.

Other than being the editor in chief of Canadian Audiologist, Marshall Chasin writes a regular column in the Hearing Review called Back to Basics. Some of these columns are reprinted in this issue of Canadian Audiologist with permission of the Hearing Review.